Drug testing devices need to be room temperature

Newsletter archive

This is an archive from the Sober Check newsletters, which are sent out every week.
Sign up to the newsletter to receive these as they come out.

Drug testing relies on reactions between antibodies (on the test strip) and antigens (in the sample, the substance being tested for).

At a lower temperature, reactions occur slower, or not at all.

If your cups are cold you will get inconsistent results, often: panels failing to run, weak/patchy lines, or colours failing to appear on the temperature strip.

Make sure your cups are warm before testing!

For storage, store your cups in a warm office or cupboard, at the temperature range the manufacturer suggests. This is usually from 2°C – 30°C.

When using a cup, the temperature range is smaller. Follow the manufacturer recommendations, usually this is from 15°C – 30°C.

Cold warehouses and vehicles can present a challenge in winter to keep cups at an ideal temperature, but a bit of planning can eliminate these concerns. Have enough cups stored in the office for a few days of testing (plus a some extras for those last minute requests!).

Cups should be stored in an insulated room overnight, as the minimum — a couple of hours in a warm office is not going to be enough to warm up really cold cups, as they’re fairly well insulated in their foil wraps.

Whoops – I ran a test and suspect the the cup was too cold. What can I do?

If you think you have an abnormal result you can test again.

If you’re using a split cup (like the Medix Pro-Split cup), the portion of the sample being tested is split off from the rest of the sample. You can pour the rest of the sample into a new, warmer cup, and repeat the test.

If you’re using a screw top cup (like the MicroScreen cup), ask the donor to give another sample.

As always, these cups are a screening test only. If in doubt after testing again, send the sample to a laboratory for confirmation.